17-62
SCPI Command Reference
:MEDian[:STATe] <b>
[:SENSe[1]]:MEDian[:STATe] <b>
Enable/disable median
fi
lter
Parameters
<b> =
0 or OFF
Disable repeat
fi
lter
1 or ON
Enable repeat
fi
lter
Query
[:STATe]?
Query state of repeat
fi
lter
Description
This command is used to enable or disable the median
fi
lter. When enabled,
voltage, current, and resistance readings are
fi
ltered according to the speci-
fi
ed rank. When disabled, the median
fi
lter stage is bypassed.
Moving
fi
lter commands
:AVERage:COUNt <n>
[:SENSe[1]]:AVERage:COUNt <n>
Set moving
fi
lter count
Parameters
<n> =
1 to 100
Specify moving
fi
lter count
DEFault
1
MINimum
1
MAXimum
100
Query
:COUNt?
Query
fi
lter count
:COUNt? DEFault
Query the *RST default
fi
lter count
:COUNt? MINimum
Query the lowest allowable
fi
lter count
:COUNt? MAXimum
Query the largest allowable
fi
lter count
Description
This command is used to specify the moving
fi
lter count. In general, the
fi
l-
ter count is the number of readings that are stored in the moving
fi
lter buffer
for the averaging calculation. After the buffer
fi
lls with readings, a new
reading goes into the buffer and the oldest reading is discarded. The
fi
lter
process if performed after each new reading is placed in the buffer.
:AVERage[:STATe] <b>
[:SENSe[1]]:AVERage[:STATe] <b>
Enable/disable moving
fi
lter
Parameters
<b> =
0 or OFF
Disable moving
fi
lter
1 or ON
Enable moving
fi
lter
Query
[:STATe]?
Query state of moving
fi
lter
Description
This command is used to enable or disable the moving
fi
lter. When enabled,
voltage, current, and resistance readings are
fi
ltered according to how the
fi
lter is con
fi
gured. When disabled, the moving
fi
lter stage is bypassed.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......