SENSe1 subsystem
The Sense1 subsystem is used to con
fi
gure and control the measurement functions of the
SourceMeter. Many of the commands are global, where a single command affects all functions.
Some commands are unique to a speci
fi
c function. For example, you can program a unique
range setting for each basic function (amps, volts, and ohms).
A function does not have to be selected before you can program its various con
fi
gurations.
Whenever a programmed function is selected, it assumes the programmed states.
The commands for this subsystem are summarized in Table 17-5.
Select measurement functions
:CONCurrent <b>
[:SENSe[1]]:FUNCtion:CONCurrent <b>
Control concurrent measurements
Parameters
<b> =
0 or OFF
Disable concurrent measurements
1 or ON
Enable concurrent measurements
Query
:CONCurrent?
Query state of concurrent measurements
Description
This command is used to enable or disable the ability of the instrument to
measure more than one function simultaneously. When enabled, the instru-
ment will measure the functions that are selected. See [:ON], :OFF and
:ALL.
When disabled, only one measurement function can be enabled. When mak-
ing the transition from :CONCurrent ON to :CONCurrent OFF, the voltage
(VOLT:DC) measurement function will be selected. All other measurement
functions will be disabled. Use the :FUNCTion[:ON] command to select
one of the other measurement functions.
[:ON] <function list>
:OFF <function list>
[:SENSe[1]]:FUNCtion[:ON] <function list>
Specify functions to be enabled
[:SENSe[1]]:FUNCtion:OFF <function list>
Specify functions to be disabled
Parameters
<function list> =
“
CURRent[:DC]
”
Amps measurement function
“
VOLTage[:DC]
”
Volts measurement function
“
RESistance
”
Ohms measurement function
NOTE
Each function in the list must be enclosed in quotes (double or single) and separated
by a comma (i.e., :func:on “volt”, “curr”).
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......