background image

 

Keithley Instruments, Inc. 

28775 Aurora Road 
Cleveland, Ohio 44139 
1-888-KEITHLEY 

www.keithley.com

 

PA-1001 Rev. D / September 2011 

13 

Model 4210-MMPC-C

Multi-Measurement Prober Cable Kit 

Quick Start Guide

Setup using the Keithley Model 4225-RPM for I-V, C-V, or pulse I-V 
measurements

 

NOTE 

The setup for three-pin pulse I-V testing requires three prober cable kits. The setups 
for four-pin testing require four prober cable kits.  

The 4225-RPM is a Remote Amplifier/Switch that is an optional item to the Model 4225-PMU. The PMU has two 
channels, so each PMU instrument card supports two (2) RPMs. The MMPC cables simplify the connection and re-
connection between the instruments and the device-under-test. The combination of the MMPC cables with the 
RPM permits testing I-V, C-V, or pulse I-V automatically, without any manual cable re-connections. When using 
only two RPMs on a four-terminal device, some re-cabling is required for certain C-V test configurations. 

The 4225-RPM has the appropriate connector types at the input and output of the RPM enclosure to eliminate any 
adapters when connecting cables to the RPM. On the input side, there are two triax connectors for SMU Force and 
SMU Sense. There are two SMA connectors for CVU POT and CVU CUR. There is also an RPM control connector 
for the control and power of the RPM from the 4225-PMU. The output of the RPM is a pair of triax cables (Force 
and Sense), which connect directly to the MMPC cables as shown in the figures below. 

See Section 16 of the 4200 Reference Manual for 4225-PMU and 4225-RPM information and how to configure an 
ITM for a pulse IV test. 

2 Channel I-V or Pulse I-V Setup 

For most I-V and pulse I-V testing, only two source/measure channels are typically used. The two active signals are 
connected to the gate and drain, and the source (and bulk, if present) is connected to ground (see Figures 17 and 
18). For the C-V test setup shown in Figure 19, all four source/measure channels are used.  

Summary of Contents for 4210-MMPC-C

Page 1: ...V A package to perform pulse I V measurements This kit has been assembled for use with the Cascade Microtech 12000 prober series manipulator type DCM 200 series For C V testing this Quick Start Guide...

Page 2: ...is a one page packing list for the Model 4210 MMPC C Multi Measurement Prober Cable Kit for the Cascade MicroTech 12000 series prober A photo shows all the components of the cable kit Section 16 of th...

Page 3: ...Figure 2 shows SMUs that do not use preamps These SMUs use the supplied triax cables that are terminated with a miniature triax connector on one end and a standard triax connector on the other end If...

Page 4: ...r Cable Kit Quick Start Guide C V testing Two prober cable kits include the cable adapters needed to connect the subminiature version A SMA cables of the Model 4210 CVU to the female triax connectors...

Page 5: ...ctors to connect the Model 4200 to the bulkhead of the prober see Figures 4 and 5 Figure 4 shows the Model 4200 connections to the prober and the trigger connection between the scope card Model 4200 S...

Page 6: ...ables and adapters used for instrumentation connections Installing the prober cable kits Basic cable setup Figure 6 shows the basic cable setup for the 2 pin I V and C V testing scenarios described in...

Page 7: ...jumpers are used to connect the commons of the cable assemblies together The jumpers plug into the grounding brackets Referring to Figure 7 connect the jumpers from one grounding bracket to another Fo...

Page 8: ...Us connected to the prober bulkhead use the fundamental setup shown in Figure 8 to perform 2 pin I V testing Figure 2 shows the SMU connections to the outside of the bulkhead Figure 8 Test setup for 2...

Page 9: ...tart Guide Figure 9 Test setup for 4 pin I V testing C V testing 2 pin C V setup NOTE The setup for 2 pin C V testing requires two prober cable kits With a Model 4210 CVU connected to the prober bulkh...

Page 10: ...test setup using the Keithley Model 4200 PIV A NOTE The setup for 4 pin pulse I V testing requires four prober cable kits The Model 4200 PIV A is a factory installed package that includes a scope car...

Page 11: ...k Start Guide NOTE Details on using the Model 4200 PIV A package are provided in the Model 4200 SCS User s Manual see How to use the Model 4200 SCS to perform a pulse I V test on my device in Section...

Page 12: ...prober cable kits The Model 4225 PMU is an Ultra Fast I V Module that is an instrument card for the Model 4200 SCS The PMU has two channels of voltage pulse source with integrated simultaneous voltag...

Page 13: ...erminal device some re cabling is required for certain C V test configurations The 4225 RPM has the appropriate connector types at the input and output of the RPM enclosure to eliminate any adapters w...

Page 14: ...Road Cleveland Ohio 44139 1 888 KEITHLEY www keithley com 14 PA 1001 Rev D September 2011 Model 4210 MMPC C Multi Measurement Prober Cable Kit Quick Start Guide Figure 17 Test setup for three pin devi...

Page 15: ...Road Cleveland Ohio 44139 1 888 KEITHLEY www keithley com PA 1001 Rev D September 2011 15 Model 4210 MMPC C Multi Measurement Prober Cable Kit Quick Start Guide Figure 18 Test setup for four pin devic...

Page 16: ...28775 Aurora Road Cleveland Ohio 44139 1 888 KEITHLEY www keithley com 16 PA 1001 Rev D September 2011 Model 4210 MMPC C Multi Measurement Prober Cable Kit Quick Start Guide Figure 19 Test setup for f...

Reviews: