Series 3700 System Switch/Multimeter Quick Start Guide
Section 4
:
Frequently Asked Questions
Document Number: 3700S-903-01 Rev. A / August 2007
4-11
Recall a setup (bus operation)
Setups are recalled over the bus from either internal memory or from a connected USB flash
drive. Recalling a setup also may be used to reset to the factory default setup. To recall a setup
over the bus, use the following command:
setup.recall
Function: Recalls
settings from a saved
setup.
setup.recall(location)
location:
Setup number to recall (0, 1, or
"/usb1/<filename>").
0: Reset setup.
1: Internal setup.
<filename>:
Use the name of the desired file
contained on a USB flash drive.
For example: Send the following to recall a user saved setup stored in a file named
KEITHLEY_3730 on an installed USB flash drive:
setup.recall("/usb1/KEITHLEY_3730")
Save/recall/reset setup (front panel)
Using the front panel, you can save the present setup to internal memory (overwriting the
existing setup), reset to the factory default settings, or recall an existing setup from internal
memory.
To save or recall a setup from the front panel:
NOTE
Press the
EXIT
key to place in local mode if the Series 3700 is in remote mode (being
controlled over the bus),
1. Press the
MENU
key (1).
2. Use the wheel (2) to select the "SETUP" menu item and press the
ENTER
key.
3. Use the wheel (2) to select menu item as desired:
SAVE: Press the
ENTER
key to have the present setup overwrite the existing internal memory
setup.
RECALL: Press the
ENTER
key
to have the internal saved setup recalled overwriting the
present setup.
RESET: Press the
ENTER
key to have the factory default setup overwrite the present setup.
NOTE
Setups saved on the USB flash drive cannot be recalled from the front panel (use the bus
to recall USB setups).