Version 1.52 Release - Non-critical fixes
Series 3700A Firmware Release Notes
3706A-850SV1.55a
/ June 2017
14
Workaround:
None.
Resolution:
The TRIGGER functionality had been implemented, but was not reported in the device and interface
capabilities. The appropriate bits in the capability words are now set and resolve this issue.
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
In the Java applet some of the file extension drop down boxes did not have a default extension
selected. This only happens when using Java JRE 7.
Workaround:
Using Java JRE 6.
Resolution:
The Java applet now requests a default file extension to be selected in the drop down box.
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
In the Java applet the Pattern and Generate Report buttons were not working.
Workaround:
None.
Resolution:
The previous firmware release inserted an issue that caused these buttons to not work properly. This
has been corrected.
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
According to VXI-11.1 rule B.4.12, the status poll (device_readstb) should automatically clear the RQS
bit, but if a second status poll is executed, the RQS bit is still set. This means the RQS bit was not
automatically clearing.
Workaround:
Manually call a status clear after a status poll.
Resolution:
The RQS bit now automatically clears after a status poll.
PR47097
PR47098
PR47099
PR47131
PR47136