Version 1.53 Release - Non-critical fixes
Series 3700A Firmware Release Notes
3706A-850SV1.55a
/ June 2017
12
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
Series 3700A web applications do not work correctly in Internet Explorer 11. Users cannot switch from
one web application to another without going through the Home page.
Workaround:
None.
Resolution:
This issue has been corrected.
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
When running Series 3700A web applications using Internet Explorer 11 with Java plug-in version 7 or
8, a warning message is shown to tell the user that "This application would like to use a version of
Java (1.6*) that is not installed on your system."
Workaround:
None.
Resolution:
This issue has been corrected.
Models affected:
Models 3706A, 3706A-S, 3706A-NFP, 3706A-SNFP.
Symptom:
USB host port issue - USB flash drives not recognized
Workaround:
None.
Resolution:
This issue has been corrected.
PR51958
PR51967
PR51000