2910-VSG Vector Signal Generator Quick Start Guide
2910-903-01 Rev. A / February 2006
21
Save and recall the instrument setup
1. Select
Menu > File > Save Setup
.
2. Select the keyboard button located next to the
File Name:
field.
3. Select the letters to spell the file name and select
Enter
.
4. Select
Save
.
5. Press the
Preset
button to return the instrument to the default setting values.
6. Select
Menu > File > Load Setup
.
7. Select (highlight) the file that you just saved.
8. Select
Load
.
Setting up a frequency and power sweep
This
example procedure
shows you how to set up a frequency sweep and a power sweep from the
front panel interface.
1. Select the
Preset
button on the front panel.
2. Select
Menu > Settings > Sweep
.
3. Select the fields in the
Sweep
dialog and enter values using one of the following methods:
Adjust the front panel puck or knob to change the value.
Double-click a value field and enter a value from the keypad, followed by the correct
terminator.
4. Select the
Start Frequency
or
Start Power
field and enter a value. This is the frequency or
power setting of the first step in the sweep.
5. Select the
Stop Frequency
or
Stop Power
field and enter a value. This is the frequency or
power setting of the last step in the sweep.
6. Select the
Number of steps
field and enter a value. (The default is 201 and the maximum is
2000.) As the source sweeps the power or frequency range, it divides the range by this
number, resulting in the number of individual instrument settings that the instrument makes
when progressing through the sweep range.
7. Select the
Dwell Time
field and enter a value. (The default is 0 seconds.) This is the time the
instrument stays at the specific settings for each step in the sweep range.
8. Select the
Sweep Now
button in the
Sweep
dialog to start the sweep.
9. Select the
Stop Sweep
button in the
Sweep
dialog to stop the sweep.
NOTE:
The main status area shows the “Sweeping” indication in both the
Frequency
and
Amplitude
fields.
Loading a digital modulation waveform
This
example procedure
shows you how to load a GSM / EDGE modulation waveform. See
“Verifying digital modulation licenses” for details.
1. Press instrument
Preset.
2. Choose the
Select Modulation
button.
3. Select
GSM / EDGE.
4. Select
Waveform File
.
5. Select
Open.