Model 2790 SourceMeter
®
Switch System User’s Manual
3-11
OPEN key
As shown in
, the OPEN:ALL option of the OPEN key opens all channels in the
mainframe (both slots), while OPEN:MULT opens multiple selected channels.
NOTE
For a 7751/7752/7753 module, the OPEN:ALL operation will not open channel
20 if the interlock is open.
Figure 3-8
OPEN key
Remote programming
The commands to control 7702 channels are listed in
.
Examples:
ROUT:OPEN:ALL
‘ Open all channels.
ROUT:CLOS (@101)
‘ Close 7702 channel 1 in slot 1.
ROUT:CLOS?
‘ Return closed channel.
Table 3-2
Commands to control 7702 module channels
Commands
Description
ROUTe:CLOSe <clist>
Specify one input channel to close.
ROUTe:OPEN <clist>
Open channel specified in list.
ROUTe:OPEN:ALL
Open all channels.
ROUTe:CLOSe?
Return closed channel.
ROUTe:CLOSe:STATe? <clist>
Query channels in specified list;
0 = open, 1 = closed.
CLOSE
OPEN
Press OPEN key
OPEN: MULT
OPEN: ALL
Display MULT option
and press ENTER
OPEN MULT:XXX
Display channel number
(XXX) and press ENTER
Display ALL option