Model 2750 Multimeter/Switch System User’s Manual
Measurement Considerations
E-7
Ground loops
When two or more instruments are connected together, care must be taken to avoid
unwanted signals caused by ground loops. Ground loops usually occur when sensitive
instrumentation is connected to other instrumentation with more than one signal return
path such as power line ground. As shown in
, the resulting ground loop causes
current to flow through the instrument LO signal leads and then back through power line
ground. This circulating current develops a small but undesirable voltage between the LO
terminals of the two instruments. This voltage will be added to the source voltage, affect-
ing the accuracy of the measurement.
Figure E-2
Power line ground loops
Signal Leads
Instrument 1
Power Line Ground
Ground
Loop
Current
Instrument 2
Instrument 3
2750-900-01.book Page 7 Wednesday, August 3, 2011 7:56 AM
Summary of Contents for 2750
Page 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Page 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Page 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Page 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Page 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...