Model 2657A High Power System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2657A-901-01 Rev. B/December 2012
7-55
digio.trigger[N].reset()
This function resets trigger values to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
digio.trigger[
N
].reset()
N
Digital I/O trigger line (1 to 14)
Details
This function resets the following attributes to factory default settings:
•
digio.trigger[
N
].mode
•
digio.trigger[
N
].pulsewidth
•
digio.trigger[
N
].stimulus
It also clears
digio.trigger[
N
].overrun
.
Example
digio.trigger[3].mode = 2
digio.trigger[3].pulsewidth = 50e-6
digio.trigger[3].stimulus = digio.trigger[5].EVENT_ID
print(digio.trigger[3].mode, digio.trigger[3].pulsewidth,
digio.trigger[3].stimulus)
digio.trigger[3].reset()
print(digio.trigger[3].mode, digio.trigger[3].pulsewidth,
digio.trigger[3].stimulus)
Set the digital I/O trigger line 3 for a falling edge with a pulsewidth of 50 microseconds.
Use digital I/O line 5 to trigger the event on line 3.
Reset the line back to factory default values.
Output before reset:
2.00 5.00000e-05 5.00
Output after reset:
0.00 1.00000e-05 0.00
Also see
(on page 7-52)
(on page 7-53)
(on page 7-54)
digio.trigger[N].stimulus
This attribute selects the event that causes a trigger to be asserted on the digital output line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Digital I/O trigger N reset
Recall setup
Not saved
0