Section 7: TSP command reference
Model 2657A High Power System SourceMeter® Instrument Reference Manual
7-52
2657A-901-01 Rev. B/December 2012
Example 2
smua.trigger.arm.stimulus =
digio.trigger[3].EVENT_ID
Uses a trigger event on digital I/O
trigger line 3 to be the trigger stimulus
for the SMU.
Also see
None
digio.trigger[N].mode
This attribute sets the mode in which the trigger event detector and the output trigger generator operate on the
given trigger line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Digital I/O trigger
N
reset
Recall setup
Not saved
0
(digio.TRIG_BYPASS)
Usage
triggerMode
= digio.trigger[
N
].mode
digio.trigger[
N
].mode =
triggerMode
triggerMode
The trigger mode; see
Details
for values
N
Digital I/O trigger line (1 to 14)
Details
Set
triggerMode
to one of the following values:
Trigger mode values
triggerMode
Description
digio.TRIG_BYPASS
or
0
Allows direct control of the line.
digio.TRIG_FALLING
or
1
Detects falling-edge triggers as input; asserts a TTL-low pulse for
output.
digio.TRIG_RISING
or
2
If the programmed state of the line is high, the
digio.TRIG_RISING
mode behavior is similar to
digio.TRIG_RISINGA
. If the programmed state of the line is low,
the
digio.TRIG_RISING
mode behavior is similar to
digio.TRIG_RISINGM
. This setting should only be used if
necessary for compatibility with other Keithley Instruments products.
digio.TRIG_EITHER
or
3
Detects rising- or falling-edge triggers as input. Asserts a TTL-low
pulse for output.
digio.TRIG_SYNCHRONOUSA
or
4
Detects the falling-edge input triggers and automatically latches and
drives the trigger line low. Asserting the output trigger releases the
latched line.
digio.TRIG_SYNCHRONOUS
or
5
Detects the falling-edge input triggers and automatically latches and
drives the trigger line low. Asserts a TTL-low pulse as an output
trigger.
digio.TRIG_SYNCHRONOUSM
or
6
Detects rising-edge triggers as input. Asserts a TTL-low pulse for
output.
digio.TRIG_RISINGA
or
7
Detects rising-edge triggers as input. Asserts a TTL-low pulse for
output.
digio.TRIG_RISINGM
or
8
Asserts a TTL-high pulse for output. Input edge detection is not
possible in this mode.
Summary of Contents for 2657A