Model 2651A High Power System SourceMeter® Instrument Reference Manual
Appendix C: LAN concepts and settings
2651A-901-01 Rev. A / March 2011
C-9
To set the IP address (when LAN configuration method is set to MANUAL):
1. From the front panel, press the
MENU
key, and then select
LAN > CONFIG > IP-ADDRESS
.
2. Turn the navigation wheel
to select and enter a valid IP address for the instrument.
3. Press
the
ENTER
key to confirm the changes.
4. Press
the
EXIT (LOCAL)
key once to return to the LAN CONFIG menu.
5. Select
APPLY_SETTINGS > YES
, then press the
ENTER
key.
Setting the gateway
NOTE
Contact your IT department to secure a valid gateway for the instrument when placing the instrument
on a corporate network.
The gateway does not need to be manually set if the LAN configuration method is set to AUTO.
To set the gateway (when LAN configuration method is set to MANUAL):
1. From the front panel, press the
MENU
key, and then select
LAN > CONFIG > GATEWAY
.
2. Turn the navigation wheel
to select and enter a valid gateway address for the instrument.
3. Press
the
ENTER
key to confirm the changes.
4. Press
the
EXIT (LOCAL)
key once to return to the LAN CONFIG menu.
5. Select
APPLY_SETTINGS > YES
, then press the
ENTER
key.
Setting the subnet mask
NOTE
Contact your IT department to secure a valid subnet mask for the instrument when placing the
instrument on a corporate network.
The subnet mask does not need to be manually set if the LAN configuration method is set to AUTO.
To set the subnet mask (when LAN configuration method is set to MANUAL):
1. From the front panel, press the
MENU
key, and then select
LAN > CONFIG > SUBNETMASK
.
2. Turn the navigation wheel
to select and enter a valid subnet mask for the instrument.
3. Press
the
ENTER
key to confirm the changes.
4. Press
the
EXIT (LOCAL)
key once to return to the LAN CONFIG menu.
5. Select
APPLY_SETTINGS > YES
, then press the
ENTER
key.