Section 2: General operation
Model 2651A High Power System SourceMeter® Instrument Reference Manual
2-6
2651A-901-01 Rev. A / March 2011
3. Power module
Contains the AC line receptacle and power line fuse. The instrument can operate
on line voltages of 100 V to 240 V AC at line frequencies of 50 Hz or 60 Hz.
4. IEEE-488
IEEE-488
Connector for IEEE-488 (GPIB) operation. Use a shielded cable, such as the
Keithley Instruments Model 7007-1 or Model 7007-2.
5. LAN
LAN
!
Use this RJ-45 connector to connect the instrument to the local area network.
This interface supports Auto-MDIX, so either a CAT-5 cross-over cable
(provided), or a normal CAT-5 straight-through cable (not provided) can be used.
6. TSP-link
!
Expansion interface that allows a Model 2651A and other TSP-enabled
instruments to trigger and communicate with each other. Use a category 5e or
higher LAN crossover cable (Keithley Instruments part number CA-180-3A).
7. Chassis ground
Ground screw for connections to chassis ground.
8. Cooling exhaust vent
Exhaust vent for the internal cooling fan. Keep the vent free of obstructions to
prevent overheating. Also see
9. Terminal connector
G
G G G
S
LO
SENSE/GUARD
S
HI
This connector provides input/output connections for sense (S HI/S LO) and
guard (G). Connections are as follows:
S LO = Sense LO
G = Guard
S HI = Sense HI
10. OUTPUT connector (40 V, 50 A maximum)
LO
HI
OUTPUT
40V, 50A MAX.
This connector provides HI/LO connection points for currents. Connections are
as follows:
LO = Input/Output LO
HI = Input/Output HI