Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-338
2606B-901-01 Rev. B / May 2018
Figure 126: SweepVLogMeasureI()
Example
SweepVLogMeasureI(smua, 1, 10,
0.001, 5)
This function performs a five-point logarithmic voltage
sweep starting at 1 V and stopping at 10 V. Current is
measured at every step (point) in the sweep after a 1 ms
source settling period.
The following table contains log values and corresponding
source levels for the five-point logarithmic sweep:
MeasurePoint LogStepSize LogStep SourceStepLevel
1
0.25
0.0
1.0000 V
2
0.25
0.25
1.7783 V
3
0.25
0.5
3.1623 V
4
0.25
0.75
5.6234 V
5
0.25
1.0
10.000 V
Also see
timer.measure.t()
This function measures the elapsed time since the timer was last reset.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
time
= timer.measure.t()
time
The elapsed time in seconds (1 µs resolution)