2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-145
Example
print(localnode.revision)
Outputs the present revision level.
Sample output:
3.3.0
Also see
(on page 7-139)
(on page 7-145)
localnode.serialno
This attribute stores the serial number of the module.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
serialno
= localnode.serialno
serialno
The serial number of the module
Details
This indicates the module serial number.
The instrument serial number is not available through a remote interface.
Example
display.clear()
display.settext(localnode.serialno)
Clears the virtual front panel display.
Places the serial number of the module on the top line of its
display.
Also see
(on page 7-139)
(on page 7-144)
localnode.showerrors
This attribute sets whether or not the instrument automatically sends generated errors.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Power cycle
Not saved
0 (disabled)
Usage
errorMode
= localnode.showerrors
localnode.showerrors =
errorMode
errorMode
Show error setting:
Show errors:
1
Do not show errors:
0