2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-139
If the
localnode.linefreq
attribute is explicitly set,
localnode.autolinefreq
is automatically
set to
false
.
When using this command from a remote node,
localnode
should be replaced with the node
reference, for example
node[5].autolinefreq
.
Also see
(on page 7-140)
localnode.description
This attribute stores a user-defined description and mDNS service name of the instrument.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile memory
Instrument specific (see
Details
)
Usage
localnode.description =
description
description
= localnode.description
description
User-defined description and mDNS service name of the instrument; use a string of
63 characters or less
Details
This attribute stores a string that contains a description of the instrument. This value appears on LXI
welcome page of the instrument. The value of this attribute is also used as the mDNS service name
of the instrument.
This attribute's factory default value is "Keithley Instruments SMU <model number> - <serial
number>", where <model number> and <serial number> are replaced with the actual model number
and serial number of the instrument. Setting this attribute to an empty string (in other words, setting
this attribute to a string of length zero, or one consisting entirely of whitespace characters) will revert
the description to the factory default value.
When using this command from a remote node,
localnode
should be replaced with the node
reference, for example
node[5].description
.
Example
description = "System in Lab 05"
localnode.description = description
Set
description
to "System in Lab 05".
Also see
None