Introduction
This appendix includes calibration programs written in BASIC to help you in calibrating the
Models 2303/2303B or Model 2303-PJ. Refer to
Section 2
for more details on calibration procedures,
equipment, and connections.
Appendix B
covers calibration commands in detail.
Computer hardware requirements
The following computer hardware is required to run the calibration programs:
• IBM PC compatible computer.
• Keithley KPC-488.2 or KPC-488.2AT, or CEC PC-488 IEEE-488 interface for the
computer.
• Two shielded IEEE-488 bus cables (Keithley Model 7007)
Software requirements
To use the calibration program, you will need the following computer software:
• Microsoft QBasic (supplied with MS-DOS 5.0 or later).
• MS-DOS version 5.0 or later.
• HP-style Universal Language Driver, CECHP.EXE (supplied with Keithley and CEC
interface cards listed above).
Calibration equipment
The following calibration equipment is required:
• Keithley Model 2001 Digital Multimeter
• 4
Ω
, 0.1%, 100W resistor
• 3k
Ω
, 0.1%, 0.25W resistor (Models 2303 and 2303B only)
• 30
Ω
, 0.1%, 10W resistor (Model 2303-PJ only)
See
Section 2
for detailed equipment specifications as well as details on test connections.
深圳德标仪器
135-1095-0799
Summary of Contents for 2303
Page 12: ...1 Performance Verification 135 1095 0799...
Page 26: ...2 Calibration 135 1095 0799...
Page 48: ...3 Disassembly 135 1095 0799...
Page 53: ...3 6 Disassembly 135 1095 0799...
Page 76: ...A Specifications 135 1095 0799...
Page 81: ...A 7 Specifications 135 1095 0799...
Page 82: ...B CalibrationReference 135 1095 0799...
Page 90: ...C CalibrationPrograms 135 1095 0799...