Table 1-5
5A range current readback accuracy limits
* As determined from digital multimeter and 1
Ω
resistor. See procedure.
Models 2303 and 2303B 5mA range readback accuracy
The following test applies only to the Models 2303 and 2303B:
1.
With the power off, connect the digital multimeter and characterized 3k
Ω
resistor to the
Model 2303 or 2303B OUTPUT SOURCE terminals, as shown in Figure 1-4. Be sure
to observe proper polarity and connections (3k
Ω
resistor between and DMM
INPUT HI; SOURCE - to DMM INPUT LO).
Figure 1-4
Connections for Model 2303 and 2303B 5mA current verification tests
Nominal output
Voltage
Model 2303
Output current*
Current readback limits
(1 Year, 18°C-28°C)
1V
1.000A
0.9976 to 1.0024A
2V
2.000A
1.9956 to 2.0044A
3V
3.000A
2.9936 to 3.0064A
4V
4.000A
3.9916 to 4.0084A
4.9V
4.900A
4.8898 to 4.9102A
NEXT
PREV
POWER
DISPLAY
2001 MULTIMETER
DCV
ACV
DCI
ACI
Ω
2
Ω
4
FREQ
TEMP
REL
TRIG
INFO
LOCAL
EXIT
ENTER
RANGE
RANGE
!
F
R
500V
PEAK
FRONT/REAR
2A 250V
AMPS
CAL
ST ORE RECALL
CHAN
SCAN
FILTER
MATH
CONFIGMENU
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1100V
PEAK
AUTO
W ARNING:
NO INTERNAL OPERAT OR SERVICABLE PAR TS,SERVICE BY QU ALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERAT OR SERVICABLE PARTS,SER VICE BY QU ALIFIED PERSONNEL ONL Y.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RA TING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
LINE RATING
100-240V AC
50, 60 HZ
185VA MAX
LINE FUSE
SLOWBLO W
2.5A, 250V
IEEE-488
(CHANGE IEEE ADDRESS
WITH FR ONT PANEL MENU)
REMO TE
DISPLAY
OPTION
ISOLATION FROM EARTH:
22 VOL TS MAX.
+
+
+
_
_
_
_
+
SOURCE
SENSE
D VM
IN
SOURCE
OUTPUT
0-20V, 0-5A
Model 2001 DMM
Model 2303
Input LO
Input HI
Source -
3k
Ω
Resistor
Note: Use 4-wire connections
to resistor terminals.
Sense -
Sense +
深圳德标仪器
135-1095-0799
Summary of Contents for 2303
Page 12: ...1 Performance Verification 135 1095 0799...
Page 26: ...2 Calibration 135 1095 0799...
Page 48: ...3 Disassembly 135 1095 0799...
Page 53: ...3 6 Disassembly 135 1095 0799...
Page 76: ...A Specifications 135 1095 0799...
Page 81: ...A 7 Specifications 135 1095 0799...
Page 82: ...B CalibrationReference 135 1095 0799...
Page 90: ...C CalibrationPrograms 135 1095 0799...