13-10
Performance Verification
2.
Select the multimeter DC voltage measuring function, and enable auto-ranging.
3.
Select Channel #1 by pressing the
▲
and
▲
keys to toggle between Channel #1 and
Channel #2. Turn on the Model 2306 output.
4.
Verify compliance current accuracy for the currents listed in
. For each test
point:
• Set the Model 2306 output voltage to 8V, and set the compliance current to the value
being tested.
• Note and record the digital multimeter voltage reading.
• Calculate the current from the voltage reading and actual 1
Ω
resistor value: I = V/R.
• Verify that the current is within the limits given in the table.
1.
Repeat steps 1 through 4 for Channel #2 (use the
▲
and
▲
keys to toggle between
Channel #1 and Channel #2).
Table 13-4
Compliance current accuracy limits
Model 2306 compliance
current setting
Compliance current limits
(2 Years, 18°C-28°C)
1.000A
0.993 to 1.007A
2.000A
1.992 to 2.008A
3.000A
2.990 to 3.010A
4.000A
3.989 to 4.011A
4.500A
4.488 to 4.512A
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Summary of Contents for 2302
Page 2: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 4: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 22: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 344: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 346: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 347: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...