
CONFIGURATION
93
Power On Configuration Settings
CV Control
Sets the constant voltage (CV) control mode
between local and external voltage/resistance
control. For external voltage control, see page
101 (External Voltage Control of Voltage
Output) and page 106 (External Resistance
Control of Voltage Output).
F-90
0= Panel control (local)
1 = External voltage control
2 = External resistance control
(Ext-R 10k
Ω
= Vo,max)
3 = External resistance control
(Ext-R
10k
Ω
= 0)
CC Control
Sets the constant current (CC) control mode
between local and external voltage/resistance
control. For details on external voltage control,
see page 104 (External Voltage Control of
Current Output) and 108 (External Resistance
Control of Current Output).
F-91
0= Panel control (local)
1 = External voltage control
2 = External resistance control
(Ext-R 10k
Ω
= Io,max)
3 = External resistance control
(Ext-R
10k
Ω
= 0)
Power-ON
Output
Sets the power supply to turn the output on or
off at power up.
F-92
0 = OFF at startup, 1 = On at
start up
Master/Slave
Sets the power supply as master or slave. See
the parallel/series operation for details, page
56.