Section 5: SCPI command reference
Series 2230G Triple-Channel Power Supplies User's Manual
5-16
2230G-900-01 Rev. A / June 2018
Example
CAL:VOLT 30.002V
Sets the voltage value of the calibration point to 30.002 V.
See also
None
CALibrate:VOLTage:LEVel
This command sets the voltage calibration points.
Group
Calibration
Syntax
CALibrate:VOLTage:LEVel <point>
Arguments
<point>
=
P1|P2|P3|P4
Calibration point 1, 2, 3, or 4.
Points P1, P2, P3, and P4 must be calibrated in numeric order.
Default value
Not applicable
Returns
Not applicable
Example
CAL:VOLT:LEV P1
Sets the voltage calibration point to point 1.
See also
None