Section 5: SCPI command reference
Series 2230G Triple-Channel Power Supplies User's Manual
5-12
2230G-900-01 Rev. A / June 2018
Returns
Not applicable
Example
CAL:CURR 3.002A
Sets the actual output current value of the calibration point to 3.002 A.
See also
None
CALibrate:CURRent:LEVel
This command sets the current calibration points.
Group
Calibration
Syntax
CALibrate:CURRent:LEVel <point>
Arguments
<point>
=
P1|P2
Point 1 or point 2
Points P1 and P2 must be calibrated in numeric order.
Default value
Not applicable
Returns
Not applicable
Example
CAL:CURR:LEV P1
Sets the current calibration point to point 1.
See also
None