Section 2: Adjustment
Model 2110 5½ Digit Multimeter Calibration Manual
2-8
2110-905-01 Rev. C / August 2013
36.
Set the Model 5700A output to 100kΩ.
37. Adjust full scale measurement by sending the following command:
CAL:PROT:DC:STEP 2,<ACTUAL CALIBRATOR OUTPUT>
38. Wait until the following message is displayed on the Model 2110:
Cali OK
39. Set the Model 5700A output mode to STANDBY.
40. Configure the Model 2110 for 2-
wire, 1MΩ resistance measurements by sending the following
command:
CONF:RES 1E6
41. Set the measurement rate to 10 PLC by sending the following command:
RES:NPLC 10
42.
Set the Model 5700A to 0Ω and the output mode to OPERATE.
43. Adjust input offset by sending the following command:
CAL:PROT:DC:STEP 1,0
44. Wait until the following message is displayed on the Model 2110:
Cali OK
45.
Set the Model 5700A output to 1 MΩ.
46. Adjust full scale measurement by sending the following command:
CAL:PROT:DC:STEP 2,<ACTUAL CALIBRATOR OUTPUT>
47. Wait until the following message is displayed on the Model 2110:
Cali OK
48. Set the Model 5700A output mode to STANDBY.
49. Configure the Model 2110 for 2-
wire, 10 MΩ resistance measurements by sending the following
command:
CONF:RES 1E7
50. Set the measurement rate to 10 PLC by sending the following command:
RES:NPLC 10
51.
Set the Model 5700A to 0Ω and the output mode to OPERATE.
52. Adjust input offset by sending the following command:
CAL:PROT:DC:STEP 1,0
53. Wait until the following message is displayed on the Model 2110:
Cali OK
54.
Set the Model 5700A output to 10MΩ.
55. Adjust full scale measurement by sending the following command:
CAL:PROT:DC:STEP 2,<ACTUAL CALIBRATOR OUTPUT>
56. Wait until the following message is displayed on the Model 2110:
Cali OK
57. Set the Model 5700A output mode to STANDBY.