Model 2110 5½ Digit Multimeter Calibration Manual
Section 1: Calibrating and adjusting the Model 2110
2110-905-01 Rev. C / August 2013
1-3
Figure 2: Command Control window
Environmental conditions
Perform the calibration procedures in an environment that has:
•
An ambient temperature of 18 °C to 28 °C
•
A relative humidity of less than 80 percent unless otherwise noted.
Calibration considerations
When performing the calibration procedure:
•
Make sure that the equipment is properly warmed up and connected to the appropriate input
jacks.
•
Make sure the calibrator is in OPERATE before you complete each calibration step.
•
Do not connect test equipment to the Model 2110 through a scanner or other switching
equipment.
•
If an error occurs during calibration, the Model 2110 will generate an appropriate error message.
Please refer to The Model 2110 Reference Manual for error code descriptions.
•
During the adjustment process, the Model 2110 may not display the correct amplitude of the
applied signal. This condition is normal.
•
After the adjustment procedures are completed, perform the performance verification in Section 3
to verify that the Model 2110 is within manufacturer’s specifications.