background image

 

 

11/07 

 

Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating 

information, and as shown on the instrument or test fixture panels, or switching card. 
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard. 
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground connections. 
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use 

of a lid interlock. 

If a 

 screw is present, connect it to safety earth ground using the wire recommended in the user documentation. 

The 

 symbol on an instrument means caution, risk of danger. The user should refer to the operating instructions located in 

the user documentation in all cases where the symbol is marked on the instrument. 

The 

 symbol on an instrument means caution, risk of electric shock. Use standard safety precautions to avoid personal 

contact with these voltages. 

The 

  symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns. 

The 

 symbol indicates a connection terminal to the equipment frame. 

If this 

 symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be 

properly disposed of according to federal, state, and local laws. 
The 

WARNING

 heading in the user documentation explains dangers that might result in personal injury or death. Always read 

the associated information very carefully before performing the indicated procedure. 
The 

CAUTION

 heading in the user documentation explains hazards that could damage the instrument. Such damage may 

invalidate the warranty. 
Instrumentation and accessories shall not be connected to humans. 
Before performing any maintenance, disconnect the line cord and all test cables. 
To maintain protection from electric shock and fire, replacement components in mains circuits — including the power 

transformer, test leads, and input jacks — must be purchased from Keithley Instruments. Standard fuses with applicable national 

safety approvals may be used if the rating and type are the same. Other components that are not safety-related may be 

purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be 

purchased only through Keithley Instruments to maintain accuracy and functionality of the product). If you are unsure about the 

applicability of a replacement component, call a Keithley Instruments office for information. 
To clean an instrument, use a damp cloth or mild, water-based cleaner. Clean the exterior of the instrument only. Do not apply 

cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with 

no case or chassis (e.g., a data acquisition board for installation into a computer) should never require cleaning if handled 

according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the 

factory for proper cleaning/servicing. 

 

Summary of Contents for 2110-100

Page 1: ...N O C F O E R U S A E M R E T A E R G A Model 2110 5 Digit Multimeter Calibration Manual 2110 905 01 Rev A July 2012 Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipm...

Page 2: ...ation herein in whole or in part without the prior written approval of Keithley Instruments Inc is strictly prohibited All Keithley Instruments product names are trademarks or registered trademarks of...

Page 3: ......

Page 4: ...ment control and data I O connections are for connection to Category I sources unless otherwise marked or described in the user documentation Exercise extreme caution when a shock hazard is present Le...

Page 5: ...dangers that might result in personal injury or death Always read the associated information very carefully before performing the indicated procedure The CAUTION heading in the user documentation expl...

Page 6: ...justment 100 M 2 11 Current function adjustment DC and AC 2 12 DC current range adjustment 2 12 DC high current adjustment 2 13 AC current range adjustment 2 14 AC high current adjustment 2 15 Tempera...

Page 7: ......

Page 8: ...proper VISA layer Keithley I O Layer 5 0 or greater included on the CD ROM that came with your Model 2110 NI VISA 3 1 or greater Agilent I O Library Suite 14 2 or greater 1 Install the USB device dri...

Page 9: ...5520A 5522A output to the Model 2110 front panel input setting up appropriate ranges and performing calibration by sending specific SCPI commands These commands are entered into the Send String field...

Page 10: ...nnected to the appropriate input jacks Make sure the calibrator is in OPERATE before you complete each calibration step Do not connect test equipment to the Model 2110 through a scanner or other switc...

Page 11: ...e Fluke 5725A Amplifier or the equivalent or better and the Fluke 5520A 5522A for temperature and capacitance to calibrate the Model 2110 Calibration code To unlock and initiate calibration 1 Unlock c...

Page 12: ...in injury or death To eliminate the thermal EMFs due to the differences between two metals use copper leads to connect your source signal to the DMM DCV range adjustment 1 Set the Model 5700A for 0 V...

Page 13: ...mode 16 Adjust the input offset by sending the following command CAL PROT DC STEP 1 0 17 Wait until the following message is displayed on the Model 2110 Cali OK 18 Set the Model 5700A output to 1 V D...

Page 14: ...2110 Cali OK 40 Set the Model 5700A output to 100 V DC 41 Adjust full scale measurement by sending the following command CAL PROT DC STEP 2 100 42 Wait until the following message is displayed on the...

Page 15: ...e Model 5700A output mode to STANDBY 11 Configure the Model 2110 for the 1 V range by sending the following command CONF VOLT AC 1 12 Set the Model 5700A to 0 1 V and the output mode to OPERATE 13 Adj...

Page 16: ...5 V and the output mode to OPERATE 37 Adjust to 10 percent of range by sending the following command CAL PROT AC STEP 1 75 38 Wait until the following message is displayed on the Model 2110 Cali OK 39...

Page 17: ...gure the Model 2110 for 2 wire 100 resistance measurements by sending the following command CONF RES 1E2 4 Set the measurement rate to 10 PLC by sending the following command RES NPLC 10 5 Disable the...

Page 18: ...Set the measurement rate to 10 PLC by sending the following command RES NPLC 10 24 Set the Model 5700A to 0 and the output mode to OPERATE 25 Adjust input offset by sending the following command CAL P...

Page 19: ...del 2110 Cali OK 45 Set the Model 5700A output to 1 M 46 Adjust full scale measurement by sending the following command CAL PROT DC STEP 2 ACTUAL CALIBRATOR OUTPUT 47 Wait until the following message...

Page 20: ...ing command CAL PROT DC STEP 1 0 8 Wait until the following message is displayed on the Model 2110 Cali OK 9 Set the Model 5700A output to 100 10 Adjust full scale measurement by sending the following...

Page 21: ...4 wire 100 k resistance measurements by sending the following command CONF FRES 1E5 32 Set the measurement rate to 10 PLC by sending the following command FRES NPLC 10 33 Set the Model 5700A to 0 and...

Page 22: ...by sending the following command CAL PROT DC STEP 2 ACTUAL CALIBRATOR OUTPUT 56 Wait until the following message is displayed on the Model 2110 Cali OK 57 Set the Model 5700A output mode to STANDBY Oh...

Page 23: ...to 10 PLC by sending the following command CURR NPLC 10 5 Set the Model 5700A output mode to OPERATE 6 Adjust front input offset by sending the following command CAL PROT DC STEP 1 0 7 Wait until the...

Page 24: ...he Model 5725 and black of the Model 2110 to black of the Model 5725 Configure the Model 5700 to output current to the Model 5725A 2 Configure the Model 2110 for the 3 A DC current range by sending th...

Page 25: ...DC STEP 2 10 26 Wait until the following message is displayed on the Model 2110 Cali OK 27 Set the Model 5700A output mode to STANDBY Press RESET on the Model 5700A front panel to reset the calibrato...

Page 26: ...8 Wait until the following message is displayed on the Model 2110 Cali OK 9 Set the Model 5700A output mode to STANDBY 10 Set the Model 5700A output to 3 A 1 kHz 11 Adjust full scale measurement by se...

Page 27: ...citance function adjustment 1 Set the Model 5520A 5522A to 1 uF range and the output mode to STANDBY 2 Using shielded low thermal EMF cables connect the Fluke 5700A output HI and LO to the Model 2110...

Page 28: ...Configure the Model 2110 for the 1 mF capacitance range by sending the following command CONF CAP 1E 3 25 Set the Model 5520A 5522A output mode to STANDBY 26 Adjust input offset by sending the follow...

Page 29: ...PROT DATE MM DD YYYY The date string must be enclosed in double quotes string Example If calibration date is 5 Apr 2012 send the following command CAL PROT DATE 04 05 2012 Calibration due date is auto...

Page 30: ...ange Test Point Accuracy Limits Upper Limit Lower Limit 100 mV 50 0000 mV 50 010 mV 49 990 mV 90 0000 mV 90 015 mV 89 985 mV 1 V 0 500000 V 0 50001 V 0 49992 V 0 900000 V 0 90001 V 0 89988 V 10 V 5 00...

Page 31: ...998 V 1 000000 V 75k Hz 1 007 V 0 993 V 1 000000 V 300 kHz 1 050 V 0 950 V 10 V 10 00000 V 10 Hz 10 02 V 9 98 V 10 00000 V 30 kHz 10 03 V 9 97 V 10 00000 V 75 kHz 10 07 V 9 93 V 10 00000 V 300 kHz 10...

Page 32: ...d reading b Max lead resistance 10 of range per lead for 100 and 1 k ranges Add 1 k per lead for all other ranges DC current accuracy verification DC CURRENT ACCURACY VERIFICATION Range Test Point Acc...

Page 33: ...099 A 2 6901 A 0 41666667 9 0000 A 60 Hz 9 0570 A 8 9430 A 0 41666667 9 0000 A 5 kHz 9 0570 A 8 9430 A Specifications valid for 2 hour warm up at 5 1 2 digits a Slow ac filter 3 Hz bandwidth b Pure si...

Page 34: ...d Measurement rate set to 10 PLC fixed Capacitance accuracy verification CAPACITANCE ACCURACY VERIFICATION Range Test Point Accuracy Limits Upper Limit Lower Limit 1 nF 1 00000 nF 1 02800 nF 0 97200 n...

Page 35: ...ge without notice All Keithley trademarks and trade names are the property of Keithley Instruments Inc All other trademarks and trade names are the property of their respective companies A G R E A T E...

Reviews: