Calibration Programs
B-4
Figure B-2
Connections for comprehensive calibration
2002 MULTIMETER
F
Input HI
Output HI
Input
LO
Output
LO
Model 2002
5700A Calibrator
Note
: Use shielded cables to minimize noise.
Enable or disable calibrator external
sense as indicated in procedure.
Sense HI
Sense LO
Sense LO
Sense HI
Figure B-3
Calibrator voltage connections
2002 MULTIMETER
Input HI
Output HI
Input
LO
Output
LO
Model 2002
5700A Calibrator
Figure B-4
Calibrator current connections
2002 MULTIMETER
Output HI
Input
LO
Output
LO
Model 2002
5700A Calibrator
Amps
Note: Be sure calibrator is set
for normal current output