:STEP7
(:CALibration:PROTected:DC:STEP7)
Purpose
To program the 10k
Ω
4-wire comprehensive calibration step.
Format
:cal:prot:dc:step7 <Cal_resistance>
Parameter
<Cal_resistance> = 9E3 to 11E3 [
Ω
]
Description
:STEP7 programs the 10k
Ω
4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from 9E3
to 11E3, but 10E3 is recommended for best results.
Example
:CAL:PROT:DC:STEP7 10E3
Program 10k
Ω
step.
:STEP8
(:CALibration:PROTected:DC:STEP8)
Purpose
To program the 100k
Ω
4-wire comprehensive calibration step.
Format
:cal:prot:dc:step8 <Cal_resistance>
Parameter
<Cal_resistance> = 90E3 to 110E3 [
Ω
]
Description
:STEP8 programs the 100k
Ω
4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from
90E3 to 110E3, but 100E3 is recommended for best results.
Example
:CAL:PROT:DC:STEP8 100E3
Program 100k
Ω
step.
:STEP9
(CALibration:PROTected:DC:STEP9)
Purpose
To program the 1M
Ω
comprehensive calibration step.
Format
:cal:prot:dc:step9 <Cal_resistance>
Parameter
<Cal_resistance> = 900E3 to 1.1E6 [
Ω
]
Description
:STEP9 programs the 1M
Ω
comprehensive calibration step. The allowable
range of the calibration resistance parameter is from 900E3 to 1.1E6. Use the
1E6 value whenever possible, or the closest possible value.
Example
:CAL:PROT:DC:STEP9 1E6
Program 1M
Ω
calibration step.
Calibration Command Reference
3-13
Summary of Contents for 2000
Page 1: ...Model 2000Multimeter Calibration Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Page 10: ...1 Performance Verification ...
Page 23: ...2 Calibration ...
Page 45: ...3 Calibration CommandReference ...
Page 65: ...A Specifications ...
Page 76: ...A 12 Specifications ...
Page 77: ...B Error Messages ...
Page 80: ...C Calibration Program ...