5.2. Flow cell
The flow cell assembly contains a pressure vessel and has been designed to withstand the
analyte conditions that are specified on your product label. For detailed technical drawings,
contact Keit.
5.2.1. High flow cell - ASM1396
The high flow cell is a ~27ml vessel that can be secured to the end of the probe for continuous
analysis of your process
. The flow cell is fitted with two ¾” NPT ports on opposing sides, in
order to accommodate a high flow rate offshoot pipe from the main process, passing the liquid
in front of the ATR. This is an alternative to installing the IRmadillo into a main pipeline.
It is fixed in place using a tri clamp that locates over the rear of the flow cell and two semi-
circular collars that fit around the probe; the clamp does not need to be tightened with
significant force. The flow cell can be rotated through 360° to accommodate the
user’s
installation.
Figure 18: High flow cell fitted to IRmadillo probe