Page 33
DOC0935 (26 July 2022 12:02)
www.keit.co.uk
Back to Table of Contents
Inspection of the sealing weld will require removal of the flange and spectrometer from the
pressure vessel and can be completed as part of regular inspection/maintenance of the
spectrometer and pipework.
Do not use impact (e.g. a hammer) to separate components if the gasket becomes stuck to
the flange and hub.
The lap-joint flange may be manufactured from Hastelloy or stainless steel. Care should be
exercised to ensure that it is not corroded by aggressive chemicals during maintenance or
installation.
5.6. Cradle - ASM1453
A mounting system for the IRmadillo spectrometer is available to simplify the installation
process. The cradle allows for easy installation of the spectrometer into a pipeline with bolting
positions for lap joint flange and hub complying with ASME/ANSI B16.5 or other flange options.
An optional ball valve is also available.
Figure 26: IRmadillo in cradle with flange and ball valve