WARRANTY 

The meter is warranted to be free for one year from the date of purchase. (Electrode and battery are not included) 

This  warranty  covers  normal  operation  and  does  not  cover  battery,  misuse,  abuse,  alteration,  tampering,  neglect, 

improper maintenance or damage resulting from leaking batteries. Proof of purchase is required for warranty repairs. 

Warranty is void if the meter has been opened. 

NOTE 

1.

 

Please operate the meter Under the guidance of professionals. 

2.

 

Soak the electrode into the distilled water or purified water for 5 – 30 minutes to activate before use. 

3.

 

Please use the standard buffer solutions for pH calibration, and operate it by the calibration steps shown above. DO 

NOT press CAL key in NON calibration mode, it may cause the calibration Error. 

4.

 

The pH buffer powder should be fully dissolved in water (250ml). 

5.

 

In  calibration  mode,  the  electrode  should  be  submerged  in  the  buffer  solution.  Especially  pay  attention  to  the 

Electrode fixed collar if it is loosen when calibration, it may cause the buffer solution enter the inside of the meter. 

How to make pH buffer solution 

There are 3 pH buffer powder supplied with the meter

 when you buy them, each powder 

should be Fully dissolved in 

water(250ml). DO NOT use the metal containers. 

Take the formulation of pH 6.8 buffer solution for example:   

1. Put the pH 6.86 buffer material (in the green sachet) in a jar (the volume should be no less than 250ml).   

2. Fill the jar with 250ml distill water. 

3. Place a glass stick in the solution, swirling it until the white powder dissolved. 

4.    Store the buffer solution in the cool and dry place. Attach a label on it for further use. 

5. For the further calibration, Ensure that there is a constant supply of fresh buffer solution in contact with the probe. 

Discard the solution after use. 

The way of the formulation of other buffer is same like that . Please mark it clear and prepare to use it for the 

future. 

The use of standard buffer

 

The buffer solutions (250ml) can be used several times by divided into portions which should be kept in a cool and dry 

place(20

-25

).  Never  reuse  the  used  portion.  The  electrode  should  be  fully  covered  by  buffer  solution  when 

calibration. 

 

 

Summary of Contents for CT-6321

Page 1: ...ectrode into the distilled water or purified water for 5 30 minutes to activate before use FEATURES Water proof and protection Data hold function for freezing the desired value Pen type digital pH Conductivity meter all in one Electrode is included easy for general purpose application LCD display with pH and temp OR Conductivity and temp Auto power off in 5 minutes Interchangeable Compact size lig...

Page 2: ... Weight 90 g included electrode OPERATING INSTRUCTION Pull out the cap before using Do not Screw it Power ON OFF Feature 1 Press ON OFF key to turn the Meter on and off 2 The Meter will shut off automatically if there is no operation in 5 minutes Self Checking Feature Executing the function automatically after the meter is power on each time 1 Press the ON OFF key to turn on the meter Self Checkin...

Page 3: ...ure compensation ATC The product is capable of measuring with automatic temperature compensation when ATC displays on the screen Interchangeable The default mode is press CAL key to switch to Calibration pH calibration Prepare for 4 measuring glasses to calibration 1 measuring glass for water and 3 measuring glasses for pH buffer solution pH4 0 pH6 8 pH9 1 Please be aware of the order of the pH bu...

Page 4: ... activated 4 press CAL key for 3 seconds until the text CAL appear on the display and release the key 5 Text SA will showing in the LCD denote that the current calibration result is saved 6 The text END will be displayed at the end The meter is ready for conductivity measurement The steps are same like that if the buffers are 1413uS cm NOTE The calibration will be failed if there is no SA on the s...

Page 5: ...fer solution Especially pay attention to the Electrode fixed collar if it is loosen when calibration it may cause the buffer solution enter the inside of the meter How to make pH buffer solution There are 3 pH buffer powder supplied with the meter when you buy them each powder should be Fully dissolved in water 250ml DO NOT use the metal containers Take the formulation of pH 6 8 buffer solution fo...

Reviews: