Text „SA‟ will showing in the LCD denote that the current calibration result is saved
6
The text „END‟ will be displayed at the end. The meter is ready for pH measurement
The steps are same like that if the buffers are pH4.0,pH7.0 and pH10.0;
Conductivity calibration
Prepare for 1 measuring glasses with 84uS/cm conductivity buffer solution, and 1 measuring glass for water.
STEPS
OPRATIONS
1
Insert the electrode into the distilled water or purified water for 10– 30 minutes to activate before press
ON/OFF key to turn the meter on.
2
Place the electrode into the measuring glass of water to clean, and wipe it up by tissue
3
Place the electrode into the 84uS/cm conductivity buffer solution, The pH value will be updated until a
stable reading is reached(The data maybe not accurate if the electrode is not be activated)
4
press CAL key for 3 seconds
until the text „CAL‟ appear on the display, and release the key
5
Text „SA‟ will showing in the LCD denote that the current calibration result is saved
6
The text „END‟ will be displayed at the end. The meter is ready for conductivity measurement
The steps are same like that if the buffers are 1413uS/cm
;
NOTE: The calibration will be failed if there is no “SA” on the screen during the operations, buffer solutions or the damage
of the electrode should be checked.
*Overange Reminder
There will be
“
1- - .-
”
on the display if the pH value is under 0 or over 14;
There will be
“
1- - .-
”
on the display if the conductivity value is under 0 or over 200; (6321)
There will be
“
1- - .-
”
on the display if the conductivity value is under 0 or over 2000; (6322)
There will be „L‟ or „H‟ on the temp. display area if the temperature is under 0
℃
or over 50
℃
.
*Low Voltage Reminder
The battery sign will be appeared on the display when battery replacement is needed.
MAINTENANCE
Keep the glass electrode clean, and cover with the protection cap when not use. Be aware of the damage of the electrode
when use it. Products should be kept and stored in a cool, dry place that away from direct sunlight.