Text „SA‟ will showing in the LCD denote that the current calibration result is saved 

The text „END‟ will be displayed at the end. The meter is ready for pH measurement 

The steps are same like that if the buffers are pH4.0,pH7.0 and pH10.0; 

 

Conductivity calibration 

Prepare for 1 measuring glasses with 84uS/cm conductivity buffer solution, and 1 measuring glass for water. 

STEPS 

OPRATIONS 

Insert the electrode into the distilled water or purified water for 10– 30 minutes to activate before press 

ON/OFF key to turn the meter on. 

Place the electrode into the measuring glass of water to clean, and wipe it up by tissue 

Place the electrode into the 84uS/cm conductivity buffer solution, The pH value will be updated until a 

stable reading is reached(The data maybe not accurate if the electrode is not be activated) 

press CAL key for 3 seconds

 

until the text „CAL‟ appear on the display, and release the key 

Text „SA‟ will showing in the LCD denote that the current calibration result is saved 

The text „END‟ will be displayed at the end. The meter is ready for conductivity measurement 

The steps are same like that if the buffers are 1413uS/cm

NOTE: The calibration will be failed if there is no “SA” on the screen during the operations, buffer solutions or the damage 

of the electrode should be checked. 

*Overange Reminder 

There will be 

1- - .-

on the display if the pH value is under 0 or over 14; 

There will be 

1- - .-

on the display if the conductivity value is under 0 or over 200;    (6321)

 

There will be 

1- - .-

on the display if the conductivity value is under 0 or over 2000;    (6322)

 

There will be    „L‟ or „H‟ on the temp. display area if the temperature is under 0

  or over 50

*Low Voltage Reminder 

The battery sign will be appeared on the display when battery replacement is needed. 

MAINTENANCE 

Keep the glass electrode clean, and cover with the protection cap when not use. Be aware of the damage of the electrode 
when use it. Products should be kept and stored in a cool, dry place that away from direct sunlight.

 

 
 
 

Summary of Contents for CT-6321

Page 1: ...ectrode into the distilled water or purified water for 5 30 minutes to activate before use FEATURES Water proof and protection Data hold function for freezing the desired value Pen type digital pH Conductivity meter all in one Electrode is included easy for general purpose application LCD display with pH and temp OR Conductivity and temp Auto power off in 5 minutes Interchangeable Compact size lig...

Page 2: ... Weight 90 g included electrode OPERATING INSTRUCTION Pull out the cap before using Do not Screw it Power ON OFF Feature 1 Press ON OFF key to turn the Meter on and off 2 The Meter will shut off automatically if there is no operation in 5 minutes Self Checking Feature Executing the function automatically after the meter is power on each time 1 Press the ON OFF key to turn on the meter Self Checkin...

Page 3: ...ure compensation ATC The product is capable of measuring with automatic temperature compensation when ATC displays on the screen Interchangeable The default mode is press CAL key to switch to Calibration pH calibration Prepare for 4 measuring glasses to calibration 1 measuring glass for water and 3 measuring glasses for pH buffer solution pH4 0 pH6 8 pH9 1 Please be aware of the order of the pH bu...

Page 4: ... activated 4 press CAL key for 3 seconds until the text CAL appear on the display and release the key 5 Text SA will showing in the LCD denote that the current calibration result is saved 6 The text END will be displayed at the end The meter is ready for conductivity measurement The steps are same like that if the buffers are 1413uS cm NOTE The calibration will be failed if there is no SA on the s...

Page 5: ...fer solution Especially pay attention to the Electrode fixed collar if it is loosen when calibration it may cause the buffer solution enter the inside of the meter How to make pH buffer solution There are 3 pH buffer powder supplied with the meter when you buy them each powder should be Fully dissolved in water 250ml DO NOT use the metal containers Take the formulation of pH 6 8 buffer solution fo...

Reviews: