100.210 - Multi Water Sampler - 12 x 50 ml - Manual
5
11
If “PLC Connected” turns red, there is no connection.
Follow these instructions:
1.
Disconnect USB cable and battery supply if
connected.
2.
Close Program on tablet.
3.
Connect USB cable to tablet and KC timer.
4.
Click the KC icon on tablet.
12
Assuming the connection is correct you are now ready
to program the individual delays.
Click on the icon “Read Settings”
13
The sampler has a short delay of 1 min. for the first
sample.
If delay for sample 1 is set to “0”
, the sampler
starts immediately after the delay.
If you estimates that, the deployment to the desired
depth takes
5 minutes you may enter the value “5”.
All remaining samples may have various delays
between every sample. You can enter an individually
delay in the range of 1 min. to 600 hours for every
sample but keep in mind, that the battery capacity
may be insufficient for very huge delays which may be
repeated several times.
14
For a sampler having 24 syringes you must click the
icon “Rosette 17
-
24” for adding the remaining
delays.
15
When programming is finished, press button “Send
New Settings
”
16
Once you are ready to deploy the sampler, start the
time lapse trigger by inserting the programming plug.