100.210 - Multi Water Sampler - 12 x 50 ml - Manual 

5

 

11 

 

If “PLC Connected” turns red, there is no connection.

 

Follow these instructions: 
 
1.

 

Disconnect USB cable and battery supply if 
connected. 

2.

 

Close Program on tablet. 

3.

 

Connect USB cable to tablet and KC timer. 

4.

 

Click the KC icon on tablet. 

 

 
 
 

12 

 
Assuming the connection is correct you are now ready 
to program the individual delays. 
 

Click on the icon “Read Settings”

 

 

 

 

 

13 

 
The sampler has a short delay of 1 min. for the first 
sample. 

If delay for sample 1 is set to “0”

, the sampler 

starts immediately after the delay.  
 
If you estimates that, the deployment to the desired 
depth takes 

5 minutes you may enter the value “5”.

 

 
All remaining samples may have various delays 
between every sample. You can enter an individually 
delay in the range of 1 min. to 600 hours for every 
sample but keep in mind, that the battery capacity 
may be insufficient for very huge delays which may be 
repeated several times. 
 

 
 
 

14 

 
For a sampler having 24 syringes you must click the 

icon “Rosette 17

-

24” for adding the remaining 

delays. 

 

 

 

 

15 

When programming is finished, press button “Send 

New Settings

 

 

 

 

 

16 

 

Once you are ready to deploy the sampler, start the 
time lapse trigger by inserting the programming plug.  

 

 

 

 
 

Summary of Contents for 100.210

Page 1: ...Multi water sampler shown with accessories optional Multi Water Sampler 12 syringes each 50 ml Model 100 210 Manual KC Denmark A S Research Equipment Limnology Oceanography Hydrobiology ...

Page 2: ...nt or to operators who ignore safety precautions or because of misuse or wrong operation Item Description 1 A complete system requires the following items 100 210 The multi water sampler 60 010 Time lapse trigger incl tablet and preinstalled software 100 217 Battery cylinder 4000 m with room for time lapse trigger or 100 226 for 6000 m 100 230 Cable from battery cylinder to motor 100 234 Plug for ...

Page 3: ...e first time set the motor moves one step forward releasing a single syringe The arm A releases the syringe by pushing the pawl to a side so the spring loaded bar pulls the syringe s piston backwards allowing a sample to be taken 4 The syringes is loaded one by one by pushing the holder downwards at the same time turn the pawl to hold the bar in its lower position 5 Three syringes have been loaded...

Page 4: ...vanced electronics preventing overcharging even if connected for a long time The Subconn connectors requires to be greased with Molykote see recommendations on page 7 10 The photo shows the programming plug to the left and a dummy for the Subconn connector optional 8 The tablet comes with pre installed software for the programming Once charged replace the charger cable with the USB cable and conne...

Page 5: ...0 the sampler starts immediately after the delay If you estimates that the deployment to the desired depth takes 5 minutes you may enter the value 5 All remaining samples may have various delays between every sample You can enter an individually delay in the range of 1 min to 600 hours for every sample but keep in mind that the battery capacity may be insufficient for very huge delays which may be...

Page 6: ...100 210 Multi Water Sampler 12 x 50 ml Manual 6 Maintenance 17 Wash all parts thoroughly with fresh water For maintenance of the Subconn connectors see pg 7 10 ...

Page 7: ...100 210 Multi Water Sampler 12 x 50 ml Manual 7 ...

Page 8: ...100 210 Multi Water Sampler 12 x 50 ml Manual 8 ...

Page 9: ...100 210 Multi Water Sampler 12 x 50 ml Manual 9 ...

Page 10: ...100 210 Multi Water Sampler 12 x 50 ml Manual 10 ...

Page 11: ...pose with special abilities and experience in this area as well as being equipped with the appropriate tools and individual safety equipment Failure to meet these requirements constitutes a risk to personal health and safety and economic damages When working on the unit in areas which are difficult to access or hazardous ensure to take adequate safety precautions for the operator and others in com...

Reviews: