Chapter 6. Calibrator Operation
34
Kaye LTR -90 Operating Manual
6.1.7b
Ramp/Soak
Table 13 shows and describes the menu selections on the Ramp/Soak Setup menu.
Figure 19: Ramp/Soak Key Sequence
Table 13: Ramp/Soak Setup Menu
Item
Function
1
NO. SETPOINTS
is the number of SETPOINTS for a Program. Eight SETPOINTs can be
set for the program. Set the maximum number of SETPOINTS necessary.
2
SOAK TIME
is the duration (in minutes) that each of the programmed SETPOINTS
temperature is maintained. The time starts when the temperature settles to within the
specified stability. The Stability Limit is set in the Temperature Setup menu (see
page 25).
3
NO. CYCLES
is the number of times that the Program is repeated.
4
DIRECTION
controls whether the SETPOINTS are sequenced in one direction “”
(ascending from 1 to 8) or in both directions
(ascending from 1 to 8 and then
descending from 8 to 1) before the sequence is repeated. If the both directions option is
selected, the Program sequences from the first SETPOINT to the last and then reverses
direction sequencing from the last to the first.
5
SETPOINTS MENU
opens a menu to set each of the SETPOINTS for the Program. Only the
number of SETPOINTS defined by NO. SETPOINTS will be displayed.
66.03°C
RAMP/SOAK SETUP
NO. SETPOINTS: 8
SOAK TIME: 15 MIN
NO. CYCLES: 1
DIRECTION:
SET
POINTS
PROG
MENU
RAMP/
SOAK
MENU
1
2
3
4
5
2
Summary of Contents for LTR -90
Page 4: ...ii no content intended for this page...
Page 14: ...Chapter 2 Specifications 8 Kaye LTR 90 Operating Manual no content intended for this page...
Page 24: ...Chapter 4 Parts and Controls 18 Kaye LTR 90 Operating Manual no content intended for this page...
Page 30: ...Chapter 5 General Operation 24 Kaye LTR 90 Operating Manual no content intended for this page...
Page 49: ......