background image

 

 

 

To make DSNU / PRNU correction, two pictures should be taken. For DSNU a reference image has 

to be recorded in dark, with lens closed (offset should be boosted) or fully removed from the camera 

and covered with a solid cap. To calibrate  for PRNU a reference image has to be recorded  with a 

uniform illumination of around 40%. These two separate steps in the flat field correction are therefore 

referred to as dark field calibration and bright field calibration, respectively. 

 

 

Figure 3 : JetCam Camera side view w/o lens 

 

3.1

 

Dark field calibration process 

The  dark  field  correction  is  the  easiest  one  to  calibrate.  It  only  requires  a  reference  image  to  be 

recorded without illumination on the image sensor. 

Follow this steps to perform dark field calibration process: 

1.

 

For dark field calibration, the light should be blocked from the sensor. This can be achieved by 

removing the lens and covering the sensor with a solid cap or closing the lens with a cap 

2.

 

Set the “Field Calibration Mode” to “Dark” (1) 

3.

 

Start camera’s stream either in free run or by applying an external trigger 

4.

 

Initiate the selected calibration with “Field Calibration Start” command Execute (2) 

5.

 

Stop camera’s stream 

6.

 

Enable the Dark filed correction (3) 

7.

 

Start camera’s stream either in free run or by applying an external trigger 

NOTE that Dark field calibration should be performed before Flat field. 

 

3

 

Using the DSNU and PRNU corrections 

Summary of Contents for JetCam

Page 1: ...ems GmbH Geisenhausenerstr 18 81379 Munich Germany 49 89 780 2970 info skyblue de www skyblue de In Great Britain Zerif Technologies Ltd Winnington House 2 Woodberry Grove Finchley London N12 0DR 44 1...

Page 2: ...RE 3 JETCAM CAMERA SIDE VIEW W O LENS 4 FIGURE 4 DARK FIELD CALIBRATION STEPS 5 FIGURE 5 FLAT FIELD CALIBRATION STEPS 6 FIGURE 6 UNIFORM LIGHT SOURCE BEFORE VS AFTER FLAT FIELD CALIBRATION 6 FIGURE 7...

Page 3: ...photo response non uniformities PRNU and or artifacts caused by the illumination and illumination optics These application notes describe how to calibrate the flat field functions and how to use them...

Page 4: ...with the socket electrical connector 3 Insert the fiber cable into the module by firmly pushing each cable into the module until you will hear a click 4 Connect the Power Adaptor to the JetCam power...

Page 5: ...brate It only requires a reference image to be recorded without illumination on the image sensor Follow this steps to perform dark field calibration process 1 For dark field calibration the light shou...

Page 6: ...optimal result Follow this steps to perform flat field calibration process 1 Prepare light source Uniform light should be applied across the sensor This can be achieved by removing the lens and settin...

Page 7: ...lable gain is not sufficient to correct for the difference between the weakest and brightest illuminated pixel segment a flat field cannot be achieved Figure 6 Uniform light source before vs after fla...

Page 8: ...arameter configurations In addition a Default UserSet is available so it will be possible to revert back to default factory settings 1 Selection between the different available user configuration Defa...

Page 9: ...and loading user set parameter configurations for a specific camera using Vision Point application Save User Set 1 Calibrate the desired camera parameters in Camera tab 2 Open User Set Control catego...

Page 10: ...Volatile memory save function call sequence 1 UserMemoryBulkErase erase all user non volatile memory 2 UserMemoryPageSelector select page 255 0 in memory 256 bytes for each page 3 UserMemoryPageAll ar...

Reviews: