- 21 -
Problem:
Error / solution
Pump does not start
•
Check connections between power supply and device
•
When mains operated, the LED (c) must light up in green
•
If you cannot solve the problem, return the device.
Pump does not suck
(does not build up
vacuum)
•
Status LED (c) lights up in red (error message)
•
Try to switch off and on the device several times. Wait a minute before
switching on again. After five unsuccessful attempts, return the device to the
manufacturer.
•
Check the suction kit:
•
Check breast shield valve (for cracks and soiling)
•
Check protective diaphragm for cracks
•
Protective diaphragm must be inserted flush and crinkle free
•
Check tube connections
The breast pump/power unit must not be opened by users/patients. For maintenance and repair
work, the device must be returned to the manufacturer.
10. Environmental protection and disposal
Please dispose of
mamivac
®
SENSITIVE-C
breast pumps in accordance with your local disposal regulations
for electrical devices.
11. Technical data
The expected service life is 5 years.
Pump mamivac
®
SENSITIVE-C:
REF 281012
Device classification:
Nominal voltage:
12 V DC
Classified acc. to 93/42/EEC,
Power consumption:
1.25 A
Appendix IX, Rule 11: Class IIa
H x W x D (mm):
190 x 112 x 145
Weight:
790 g
Max. vacuum:
-33 kPa (±15 %)
Mean vacuum:
Low flow intermittent sucking without continuous vacuum
Power unit (24 W):
REF 283008 EU version
(exchangeable plug-in adapters available on request)
Nominal input voltage:
100-240 V AC 50 / 60 Hz
Input current:
max. 550 mA
Output voltage:
12 V DC
Output current:
2000 mA
Protection class:
II
H x W x D (mm):
110 x 40 x 80
Weight:
186 g
12. Conversion table of units
1 kPa
corresponds:
0.01 bar
0.14504 PSI
corresponds:
7.5006 mmHg
(pounds per square inch)
(millimetre mercury, at 0 °C)