KaVo SONICflex seal 45 Instructions For Use Manual Download Page 6

Instructions for use SONICflex seal 

1 User instructions

6 / 20

Hazard levels

The warning and safety notes in this document must be observed to prevent
personal injury and material damage. The warning notes are designated as
shown below:

 HAZARD

In cases which – if not prevented – directly lead to death or severe injury.

 WARNING

In cases which – if not prevented – can lead to death or severe injury.

 CAUTION

In cases which – if not prevented – can lead to minor or moderate injury.

CAUTION

In cases which – if not prevented – can lead to material damage.

Summary of Contents for SONICflex seal 45

Page 1: ...Instructions for use SONICflex seal tip no 45 1 000 8323 SONICflex seal tip no 45 A 1 007 1506 SONICflex seal...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...nd storage conditions 10 4 Startup and shut down 11 4 1 Inserting the SONICflex tips 11 4 2 Removing the SONICflex tip 12 5 Operation 13 5 1 Setting the power on the SONICflex 13 5 2 Notes for applica...

Page 4: ...KaVo Technical Service If you have any technical questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavo com Target group The instructions for use...

Page 5: ...ccompanying documents Follow the electronic instructions for use HIBC Code UDI symbol CE mark for medical devices Medical device labelling of medical devices Transportation and storage conditions temp...

Page 6: ...nd material damage The warning notes are designated as shown below HAZARD In cases which if not prevented directly lead to death or severe injury WARNING In cases which if not prevented can lead to de...

Page 7: ...ury to soft tissue do not continue treatment in the oral cavity with instruments driven by compressed air 4 Place the torque wrench on the tip when the handpiece is in its holder 2 2 Technical conditi...

Page 8: ...ccessories that have been approved for combination with the product by the manufacturer 4 Do not make any modifications to the device unless these have been ap proved by the manufacturer of the produc...

Page 9: ...cal device is to be used exclusively for the described application while complying with The applicable occupational safety regulations The applicable accident prevention regulations And to use these i...

Page 10: ...rt thread Tips with A can be used for all versions of KaVo SONICflex quick 2008 3 5 Transportation and storage conditions CAUTION Startup after refrigerated storage Malfunction 4 Prior to startup stro...

Page 11: ...to 6 Reprocessing steps in accordance with ISO 17664 1 ISO 17664 2 Page 14 Current packaging law Dispose of and recycle the packaging appropriately in accordance with current packaging law employing...

Page 12: ...turning it clockwise 4 Use the thin region of the handle for insertion 4 Use the thick region of the handle for tightening or detachment NOTE The tip is properly tightened when the torque wrench skip...

Page 13: ...the table LEVEL 1 LEVEL 2 LEVEL 3 brief CAUTION Hazard from damage to the enamel Any extended use of level 3 is associated with the risk of damage to the enamel and premature tip breakage 4 Level 3 sh...

Page 14: ...tment 4 Remove all residual cement composite or blood immediately 4 Wipe disinfect the medical device before transport 4 Remove the tip from the medical device 4 Do not place in solutions or similar s...

Page 15: ...damage 4 Reprocess it in a washer disinfector only CAUTION Never reprocess this medical device in an ultrasonic cleaner Malfunction and material damage 4 Reprocess it in a washer disinfector only 6 3...

Page 16: ...he washer disinfector 6 3 3 Automated drying The drying procedure is usually part of the cleaning programme of the washer disinfector NOTE Please comply with the instructions for use of the washer dis...

Page 17: ...F 7 4 o F Steriliser using the gravity method at least 10 minutes at 134 o C 1 o C 4 o C 273 o F 1 6 o F 7 4 o F at least 30 minutes at 121 o C 1 o C 4 o C 250 o F 1 6 o F 7 4 o F 4 Remove the medica...

Page 18: ...specialised dental dealers Consumables Material summary Material number Torque wrench 1 000 4887 Nozzle needle 0 410 0911 Cleaning adapter Material summary Material number Cleaning adapter for SONICfl...

Page 19: ......

Page 20: ...1 000 8325 s 20240313 03 en...