Instructions for use SONICflex seal
6 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 6.4 Care products and systems - Servicing
16 / 20
4
Use the torque wrench to screw the SONICflex tip
②
to cleaning adapter
①
in a clockwise direction.
4
Remove the torque wrench.
4
Attach cleaning adapter
①
to Miele adapter AUF 1 in combination with ADS
1.
KaVo recommends washer disinfectors in accordance with EN ISO 15883-1,
which are operated using alkaline cleaning agents.
The validations were performed using a Miele washer disinfector and the
"VARIO-TD" programme and a mild alkaline cleaner made by Dr. Weigert.
In addition, KaVo recommends the use of a rinsing agent.
4
For programme settings as well as cleansers and disinfectants, please refer
to the Instructions for Use of the washer disinfector.
4
For adaptations, unless otherwise specified, refer to the Instructions for Use
of the washer disinfector.
6.3.3 Automated drying
The drying procedure is usually part of the cleaning programme of the washer
disinfector.
NOTE
Please comply with the instructions for use of the washer disinfector.
4
In order not to affect the KaVo medical device, make sure that the product
is dry on the inside and outside after completion of the cycle.
6.4 Care products and systems - Servicing
Not applicable.
6.5 Packaging
NOTE
The sterile goods package must be large enough to accommodate the product
without stretching the packaging. The quality and use of the packaging of the
items to be sterilised must meet the applicable standards and be appropriate
for the sterilisation process!
4
Seal the medical device separately in a sterile pack.
6.6 Sterilisation
Sterilisation in a steam steriliser (autoclave) in
accordance with EN 13060 / EN ISO 17665-1
CAUTION
Contact corrosion due to moisture.
Damage to product.
4
Immediately remove the product from the steam steriliser after the steril-
isation cycle.
The medical device has a max. temperature resistance of up to 138
o
C (280.4
o
F).