CAUTION
Premature wear and malfunctioning from improper storage during long periods
of nonuse.
Reduced product life.
▶ The medical device should be cleaned, serviced and stored in a dry location,
according to instructions, before long periods of nonuse.
CAUTION
Hazard from using non-KaVo equipment.
This can damage the product and cause malfunction.
▶ The tips may not be used on non-KaVo equipment.
CAUTION
The tip may release from the SONICflex during treatment.
This can endanger the care provider and patient.
▶ Check the tip for secure attachment before treatment.
CAUTION
Hazard from a bent tip due to frequent use or dropping the handpiece.
SONICflex tips can break or become contaminated.
▶ We recommend replacing the SONICflex tips every 9 to 12 months.
CAUTION
Risk of injury and infection when changing the SONICflex tips.
This can substantially endanger the user.
▶ Use gloves or finger stalls whenever you check, insert or remove the
SONICflex tips.
CAUTION
The SONICflex tips may break if the wrong or excessive output is used.
This can endanger the user and patient.
▶ Do not use the wrong or excessive output setting.
Instructions for use For SONICflex tips prep crown round, prep crown round A - REF 1.008.6383,
1.008.6384, prep crown plain, prep crown plain A - REF 1.008.6385, 1.008.6386