The KaVo medical device has a maximum temperature resistance up to 138 ℃
(280.4 °F).
(Depending on the available autoclave,) select a suitable procedure from the fol‐
lowing sterilisation processes:
▪ Autoclave with three times initial vacuum:
– at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
▪ Autoclave using the gravitation method:
– at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F) or
alternatively
– at least 60 minutes at 121 °C -1 °C / +4 °C (250 °F -1.6 °F / +7.4 °F)
▶ Use according to the manufacturer's Instructions for Use.
6.8
Storage
▶ Reprocessed products should be stored protected from dust with minimum ex‐
posure to germs in a dry, dark and cool space.
▶ Comply with the expiry date of the sterilised items.
Instructions for use For SONICflex tips prep crown round, prep crown round A - REF 1.008.6383,
1.008.6384, prep crown plain, prep crown plain A - REF 1.008.6385, 1.008.6386
6 Reprocessing methods according to ISO 17664 | 6.8 Storage