Instructions for use SMARTmatic PROPHY S31 - 1.011.6760 SMARTmatic PROPHY S31 K - 1.011.6761 SMARTmatic
PROPHY S33 - 1.011.6800
7 Reprocessing steps in accordance with ISO 17664 | 7.4 Drying
7.4 Drying
Manual Drying
▶
Blow off the outside and inside with compressed air until water drops are no
longer visible.
Automatic Drying
The drying procedure is normally part of the cleaning program of the thermod-
isinfector.
▶
Please observe the instructions for use of the thermodisinfector.
7.5 Care products and systems - Servicing
NOTICE
Improper service and care.
Premature wear and reduced product service life.
▶
Service regularly with suitable agents.
Note
Remove the insert for care.
Note
KaVo guarantees the proper function of KaVo products only if the care
products listed as accessories are used, since these were tested for proper
use on our products.
7.5.1 Servicing with KaVo Spray
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.
▶
Take screw-in and/or snap-on polishing attachments off the contra-angle
handpiece.
▶
Cover the product with the Cleanpac bag.
▶
Plug the product onto the cannula, and press the spray button for one
second.
7.5.2 Servicing with KaVo SPRAYrotor
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.
▶
Take screw-in and/or snap-on polishing attachments off the contra-angle
handpiece.
▶
Place the product on the appropriate coupling on the KaVo SPRAYrotor and
cover it with the Cleanpac bag.
▶
Service the product.
19 / 26