6 Troubleshooting | 6.1 Troubleshooting
6 Troubleshooting
CAUTION
Missing or damaged O-rings.
Malfunction and premature failure.
▶
Make sure that all O-rings are on the coupling and are undamaged.
CAUTION
Heating of the product.
Burns or product damage from overheating.
▶
Do not use the product if it is irregularly heated.
▶
The medical device is too hot while working:
Service the medical device.
▶
When the speed drops or is uneven:
Service the medical device.
▶
An O-ring is missing on the motor coupling:
Replace O-ring.
See also:
2
Instructions for use of motor
6.1 Troubleshooting
6.1.1 Replacing the O-rings
CAUTION
Hazard from improper care of the O-rings.
Functional problem or complete failure of the medical device.
▶
Do not use Vaseline or other grease or oil.
Note
The O-rings on the coupling may only be lubricated with a cotton ball wetted
with KaVo Spray.
▶
Press the O-ring between your fingers to form a loop.
▶
Push the O-ring to the front, and remove it.
▶
Insert new O-rings into the grooves.
15 / 26