Instructions for use PHYSIO Evo/Evo F/One
5 Disassembly | 5.3 Dismantling the gas-pressure spring
24 / 40
5.3 Dismantling the gas-pressure spring
CAUTION
Damage due to working on the pressure vessel.
Forceful opening of the pressure vessel or extensive heat exposure can lead to
injury and damage.
4
Only instructed specialists may work on or exchange parts in the vicinity of
the seat height adjusting element.
5.3.1 Dismantling the gas-pressure spring Evo / One
NOTE
If the gas-pressure spring is knocked out of the stool base too violently, this
could lead to damage to the paintwork of the dentist chair PHYSIO Evo/One.
To prevent damage to the paintwork, you can place a piece of wood between
the gas-pressure spring and the hammer.
4
To prevent damage to the paintwork of the operator stool, carefully hammer
the gas-pressure spring out of the stool base using a large rubber hammer.
You can place a piece of wood between the gas-pressure spring and the
hammer.
4
Hammer the gas-pressure spring out of the stool base as far as the stool
support.