![KaVo MASTERtorque Mini LUX M8700 L Instructions For Use Manual Download Page 10](http://html2.mh-extra.com/html/kavo/mastertorque-mini-lux-m8700-l/mastertorque-mini-lux-m8700-l_instructions-for-use-manual_5096420010.webp)
Instructions for use MASTERtorque Mini LUX M8700 L - 3.001.0000MASTERtorque Mini LUX M8700 LK -
3.001.3400MASTERtorque Mini LUX M8700 LS - 3.001.0600
3 Product description | 3.2 Technical Specifications M8700 L / M8700 LK
10 / 30
3.2 Technical Specifications M8700 L / M8700 LK
Drive pressure
2.1 to 3.5 bar (30 to 51 psi)
Recommended drive pressure
2.8 bar (41 psi)
Return air pressure
< 0.5 bar (7 psi)
Spray water pressure
0.8 to 2.0 bar (12 to 29 psi)
Spray air pressure
1.0 to 2.0 bar (15 to 29 psi)
Air consumption
44 to 51 Nl/min
Idle speed
380,000 to 450,000 min
-1
Recommended pressure
2 to 3 N
Attachable to all MULTIflex (LUX) / MULTIflex LED couplings.
3.3 Technical Specifications M8700 LS
Drive pressure
2.6 to 3.0 bar (38 to 44 psi)
Recommended drive pressure
2.8 bar (41 psi)
Return air pressure
< 0.5 bar (7 psi)
Spray water pressure
0.8 to 2.0 bar (12 to 29 psi)
Spray air pressure
1.0 to 2.0 bar (15 to 29 psi)
Air consumption
40 to 48 Nl/min
Idle speed
380,000 to 450,000 min
-1
Recommended pressure
2 to 3 N
Can be plugged onto all Sirona quick couplings.
3.4 Transportation and storage conditions
NOTICE
Startup after refrigerated storage.
Malfunction.
▶
Prior to startup, strongly refrigerated products must be allowed to warm up
to a temperature of 20 °C to 25 °C (68 °F to 77 °F).
Temperature: -20°C to +70°C (-4°F to +158°F)
Relative humidity: 5% RH to 95% RH absence of condensation
Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)
Protect from moisture