Instructions for use

MASTERtorque Mini LUX M8700 L - 3.001.0000
MASTERtorque Mini LUX M8700 LK -
3.001.3400
MASTERtorque Mini LUX M8700 LS -
3.001.0600

Summary of Contents for MASTERtorque Mini LUX M8700 L

Page 1: ...Instructions for use MASTERtorque Mini LUX M8700 L 3 001 0000 MASTERtorque Mini LUX M8700 LK 3 001 3400 MASTERtorque Mini LUX M8700 LS 3 001 0600...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...ng tool or diamond grinder 15 6 Troubleshooting 16 6 1 Changing the O rings on the MULTIflex LUX MULTIflex LED coupling M8700 L 16 6 2 Cleaning the spray nozzle 16 6 3 Changing the water filter 16 7 R...

Page 4: ...ge 25 8 Optional aids 26 9 Terms and conditions of warranty 27 Instructions for use MASTERtorque Mini LUX M8700 L 3 001 0000MASTERtorque Mini LUX M8700 LK 3 001 3400MASTERtorque Mini LUX M8700 LS 3 00...

Page 5: ...ce For repairs please contact your local dealer or the KaVo Repair Service directly 49 0 7351 56 1900 service reparatur kavokerr com Target group This document is intended for dentists and their assis...

Page 6: ...nditions Air pressure Transportation and storage conditions Humidity Protect from moisture Protect from impact HIBC Code Hazard levels The warning and safety notes in this document must be observed to...

Page 7: ...ust be appropriately repro cessed or sterilised In the case of injury to soft tissue do not continue treatment in the oral cavity with compressed air driven instruments 2 2 Technical condition A damag...

Page 8: ...juries Never touch soft tissue with the handpiece head or instrument cover Do not use the medical device as a light probe Use an appropriate light probe for illumination of the oral cavity or site of...

Page 9: ...he following applications Removal of carious material cavity and crown preparations removal of fillings processing of tooth and restora tion surfaces A medical device according to relevant national st...

Page 10: ...X MULTIflex LED couplings 3 3 Technical Specifications M8700 LS Drive pressure 2 6 to 3 0 bar 38 to 44 psi Recommended drive pressure 2 8 bar 41 psi Return air pressure 0 5 bar 7 psi Spray water press...

Page 11: ...rding to EN ISO 7494 2 4 1 Mounting the MULTIflex LUX MULTIflex LED coupling to M8700 L WARNING Detachment of the medical device during treatment A medical device that is not properly locked in place...

Page 12: ...coupling and the medical device and check the following pressures Drive pressure recommended drive pressure return air pressure spray water pressure and spray air pressure See also 2 3 2 Technical Sp...

Page 13: ...ts service life Make sure that the turbine is accurately coupled and firmly seated on the coupling Mount the medical device accurately on the MULTIflex LUX MULTIflex LED coupling and push it backward...

Page 14: ...tment Never use dental burs or diamond grinders with damaged or worn shafts CAUTION Danger of injury from cutters or grinders Infections or cuts Wear gloves or finger stalls CAUTION Hazard from defect...

Page 15: ...utton while the dental bur or diamond grinder is ro tating Do not touch the dental bur or diamond grinder while it is rotating Never touch soft tissue with the handpiece head or instrument cover Remov...

Page 16: ...between your fingers to form a loop Push the O ring to the front and remove it Insert new O rings into the grooves 6 2 Cleaning the spray nozzle WARNING Hazard from nonsterile products Infection dang...

Page 17: ...overheat and damage the tooth Check the filter and exchange if necessary Check the spray water channels and clean the spray nozzles with the nozzle needle Mat no 0 410 0921 if necessary Unscrew the s...

Page 18: ...essing To minimise the risk of infection during reprocessing always wear protect ive gloves Remove the tool from the medical device Remove all residual cement composite or blood immediately Do not pla...

Page 19: ...ng can only be carried out with washer disin fectors in accordance with EN ISO 15883 1 7 2 3 Manual external disinfection WARNING Incomplete disinfection Infection hazard Only use disinfection procedu...

Page 20: ...atibility of materials e g WL cid made by ALPRO Cover the medical device with the KaVo CLEANpac bag and place it on the corresponding care adapter Hold the can vertically Press the spray key for at le...

Page 21: ...cess this medical device in an ultrasonic device Malfunction and material damage Clean manually or in a washer disinfector only 7 3 1 Automated internal and external cleaning and internal and external...

Page 22: ...Spray KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation Remove the tool from the medical device Cover...

Page 23: ...e PLUS and attach it to coupling service point four on the far right A MULTIflex adaptor must be mounted there Press the handpiece together with the guide bush of the chuck to be ser viced against the...

Page 24: ...as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation Remove the tool from the medical device Service the product in the QUATTROcare See also 2...

Page 25: ...p to 138 280 4 F Sterilisation parameters Select a suitable procedure depending on the available autoclave from the fol lowing sterilisation processes Steriliser with triple pre vacuum at least 3 minu...

Page 26: ...Wrench for water filter 1 002 0321 Instrument stand 2151 0 411 9501 Nozzle pin 0 410 0921 Cleanpac 10 units 0 411 9691 MULTIflex spray head nozzle 0 411 9921 Spray head SIRONA for KaVo Spray 1 005 836...

Page 27: ...ry KaVo shall not be liable for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or co...