7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use
7 Reprocessing steps in accordance with ISO 17664
7.1 Preparations at the site of use
WARNING
Hazard from nonsterile products.
There is a risk of infection from contaminated medical devices.
▶ Take suitable personal protective measures.
Note
For preparation, remove the head from the shank.
▶ Remove all residual cement, composite or blood without delay.
▶ Reprocess the medical device as soon as possible after treatment.
▶ The medical device must be dry when transported for reconditioning.
▶ Do not place it in a solution or similar.
7.2 Cleaning
CAUTION
Malfunctions from cleaning in the ultrasonic unit.
Defects in the product.
▶ Only clean manually or in a thermodisinfector.
7.2.1 Manual cleaning - external
Accessories required:
▪ Tap water 30
o
C ± 5
o
C (86
o
F ± 10
o
F)
▪ Brush, e.g. medium-hard toothbrush
▶ Brush off under flowing tap water.
7.2.2 Automated external cleaning
KaVo recommends washer disinfectors in compliance with EN ISO 15883-1, which
are operated using alkaline cleaning agents.
▶ For program settings as well as cleansers and disinfectants to be used, please re‐
fer to the Instructions for Use of the thermodisinfector.
▶ In order to prevent negative effects on the medical device, make sure that the in‐
terior and the exterior of the medical device are dry after completion of the cycle,
and then grease it immediately with servicing agents from the KaVo care system.
7.2.3 Manual cleaning - internal
Validated interior cleaning (residual protein removal) can only be accomplished with
KaVo CLEANspray and KaVo DRYspray.
15 / 24