Instructions for use
MASTERmatic LUX M07 L - 1.009.3610
MASTERmatic LUX M20 L - 1.009.3620
MASTERmatic LUX M29 L - 1.009.3580
Page 1: ...Instructions for use MASTERmatic LUX M07 L 1 009 3610 MASTERmatic LUX M20 L 1 009 3620 MASTERmatic LUX M29 L 1 009 3580...
Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 D 88400 Biberach www kavo com...
Page 3: ...ter filter 14 7 Reprocessing steps in accordance with ISO 17664 15 7 1 Preparations at the site of use 15 7 2 Cleaning 15 7 2 1 Manual cleaning external 15 7 2 2 Automated external cleaning 15 7 2 3 M...
Page 4: ...Tools and consumables 21 9 Warranty terms and conditions 22 Instructions for use MASTERmatic LUX M07 L 1 009 3610 MASTERmatic LUX M20 L 1 009 3620 MASTERmatic LUX M29 L 1 009 3580 Table of contents 4...
Page 5: ...ly Copyright by KaVo Dental GmbH Symbols Refer to the chapter on Safety Warning symbol Important information for users and service technicians Can be steam sterilised at 134 o C 1 o C 4 o C 273 o F 1...
Page 6: ...r will help avert property damage and injury CAUTION CAUTION indicates a hazardous situation that can cause damage to property or mild to moder ate injuries WARNING WARNING indicates a hazardous situa...
Page 7: ...ac cording to instructions before long periods of nonuse The following individuals are authorized to repair and service KaVo products Technicians at KaVo branches throughout the world Technicians spec...
Page 8: ...ERmatic LUX M29 L 1 009 3580 3 Product description 3 1 Purpose Intended use 8 24 3 Product description MASTERmatic LUX M07 L Mat no 1 009 3610 MASTERmatic LUX M20 L Mat no 1 009 3620 MASTERmatic LUX M...
Page 9: ...tion by a knowledgeable user You need to comply with the following the applicable health and safety regulations the applicable accident prevention regulations these Instructions for use According to t...
Page 10: ...peed max 40 000 rpm Step down ratio 7 4 1 Maximum speed max 5 000 min 1 identification 2 green rings Spray water pressure 0 8 to 2 0 bar 12 to 29 psi Spray air pressure 1 0 to 2 0 bar 15 to 29 psi Spr...
Page 11: ...he product in the appropriate manner Prior to disposal the product and accessories must be appropriately prepared or sterilised if this is necessary 4 1 Check the amount of water CAUTION Overheating o...
Page 12: ...the LUX motor coupling and turn it until the latch audi bly snaps into place Pull on the medical device to make sure that it is securely affixed to the coupling 5 2 Remove the medical device Unlock th...
Page 13: ...ops or is uneven Caring for the medical device An O ring is missing on the motor coupling Replace O ring 6 2 Troubleshooting 6 2 1 Replacing the O rings CAUTION Hazard from improper care of the O ring...
Page 14: ...3 Changing the water filter CAUTION Hazard from insufficient spray water Insufficient spray water can cause the medical device to overheat and damage the tooth Check the filter and exchange if necessa...
Page 15: ...ns from cleaning in the ultrasonic unit Defects in the product Only clean manually or in a thermodisinfector 7 2 1 Manual cleaning external Accessories required Tap water 30 o C 5 o C 86 o F 10 o F Br...
Page 16: ...Vo recommends washer disinfectors in compliance with EN ISO 15883 1 which are operated using alkaline cleaning agents For programme settings as well as cleansers and disinfectants to be used please re...
Page 17: ...that have been released by KaVo with respect to the compatibility of materials e g WL cid made by ALPRO Blow off with compressed air until no water drops are visible Immediately after internal disinfe...
Page 18: ...icing the product as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation Cover the product with the Cleanpac bag Plug the product onto the cannul...
Page 19: ...t See also 2 Instructions for use KaVo QUATTROcare 2104 2104A 7 5 4 Servicing with KaVo QUATTROcare PLUS 2124 A Note The shank can be serviced either alone or with a head attached to it Servicing and...
Page 20: ...care Reduced product life Before each sterilisation cycle service the medical device with KaVo care prod ucts CAUTION Contact corrosion due to moisture Damage to product Immediately remove the product...
Page 21: ...ers Material summary Mat No Instrument stand 2151 0 411 9501 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Nozzle pin 0 410 0921 Replacement filter 1 002 0271 Wrench 1 002 0321 O rin...
Page 22: ...held liable for defects and their consequences that have arisen or may arise from to natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection inst...
Page 23: ......
Page 24: ...1 010 9104 bd 20211130 03 en...