![KaVo MASTERmatic LUX M07 L - 1.009.3610 Instructions For Use Manual Download Page 66](http://html1.mh-extra.com/html/kavo/mastermatic-lux-m07-l-1-009-3610/mastermatic-lux-m07-l-1-009-3610_instructions-for-use-manual_3928057066.webp)
CAUTION
Contact corrosion due to moisture.
Damage to product.
▶ Immediately remove the product from the steam steriliser after the
sterilisation cycle!
The KaVo medical device has a maximum temperature resistance of up to
138 ℃ (280.4 °F).
Select a suitable procedure (depending on the available autoclave) from
the following sterilisation processes:
▪ Autoclave with three times pre-vacuum:
- at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4 °F)
7 Reprocessing steps in accordance with ISO 17664
66 / 74
Summary of Contents for MASTERmatic LUX M07 L - 1.009.3610
Page 16: ...3 Product description MASTERmatic LUX M07 L Mat no 1 009 3610 3 Product description 16 74...
Page 17: ...MASTERmatic LUX M20 L Mat no 1 009 3620 3 Product description 17 74...
Page 18: ...MASTERmatic LUX M29 L Mat no 1 009 3580 3 Product description 18 74...
Page 29: ...4 Start up and shut down 29 74...
Page 42: ......
Page 43: ......
Page 72: ...9 Warranty terms and conditions 72 74...
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ...1 010 9104 kb 20150211 02 en...