8 Faults
|
8.4 Troubleshooting
Malfunction
Cause
Remedy
The entrance doors to the treatment
room are provided with door swit‐
ches and are opened.
▶ Close the entrance doors.
No door switches are installed and
the jumper for the door contact con‐
nection is not correctly inserted.
▶ Correctly insert and screw down
the jumper supplied
(Mat. no. 0.236.2394).
See also: 9.7 Mounting of door swit‐
92/99