4.16.5
Complete detector adjustment
The following information appears on the touchscreen: "The detector adjustment
was successfully completed!"
Pressing the OK key saves the result and returns you to the start menu.
Pressing the button "Repeat detector adjustment" returns you to step 1 of the de‐
tector adjustment menu.
4.17
Spray regulation
The amount of spray water has a significant influence on the effectiveness of laser
treatment. When using the handpiece 2060, adjust the amount of spray water so
that a thin film of water forms upon the sprayed surface. The flow of water and
suction must be adjusted to the laser parameters. When the flow of water lessens,
clean the channels with the nozzle needle.
See also: Handpiece instructions
Note
Make sure that the laser hose is inserted and screwed tight in the coupling, and
that a handpiece is connected.
Make sure that the KEY Laser 3+ / III upgraded is switched to stand by.
▶ Remove the handpiece from the holder.
▶ Hold down the key to release the spray.
The amount of spray and water is controlled by turning the scale nut.
Instructions for use KEY Laser 3+ 1343 KEY Laser III 1243 upgraded
4 Operation | 4.17 Spray regulation
55/76