Press the PAGE UP/PAGE DOWN keys to move the selection bar six lines up or
down.
Press the OK key to accept the setting that has been made and return to the higher-
level menu.
Press the Stop key to return to the higher-level menu without accepting the selection
that has been made.
Press the S-pulse key to activate the short pulse.
4.6
Ready state
CAUTION
Laser beam hazard
Irreversible eye damage.
▶ Observe the instructions accompanying the laser safety goggles.
▶ Do not activate the laser when the handpiece is not mounted.
▶ Never look into the distal end of the handpiece, laser hose coupling or the laser
hose, either with or without the laser safety goggles!
Press the READY key in the start menu or the corresponding key on the foot control
to switch the KEY Laser 3+ / III upgraded from standby to ready.
The KEY Laser 3+ / III is in ready mode when the READY key is bright with dark
lettering, and the red pilot laser exits the front of the handpiece. Laser pulses can
only be emitted in this mode.
Instructions for use KEY Laser 3+ 1343 KEY Laser III 1243 upgraded
4 Operation | 4.6 Ready state
37/76